How to Read a Probe Calibration Report

WINDTUNER

25  Nov  2026

Ethernet Intelligent Pressure Scanners
When you receive a probe calibration report, find the probe it covers, the conditions under which it was calibrated, and the tables or curves containing the results. Those are the pages to compare with your test plan. For a WINDTUNER pneumatic probe calibration report, the calibration conditions table is a good place to start.
 

Find the results by probe ID

One report may cover several probes submitted together. They may have the same name and shape and may even have been calibrated under the same conditions. Use the probe ID to find the right results. Check the ID on the probe or in the submission record against the report. Then find the corresponding tables and curves in the calibration conditions table. If several figures have similar titles, note the table and figure numbers and check them again when you turn to those pages.

If you swap in a probe of the same shape after testing has started, update the probe ID in the test log and use the calibration data for the replacement. Check the ID even if the fixture position and connections are the same. That way, anyone reviewing the results can tell which probe was in use.
Probe Calibration

Compare the planned test with the calibration conditions

The report may list airspeed or Mach number. For probes used to measure flow direction, also check the angles covered by the calibration. A report for a three-hole probe might list one calibration speed and a set of flow angles. A five-hole probe report might list angles in two directions at several Mach numbers. Read the speed and angle entries together to see whether they cover the conditions you want to test.

Look at the actual calibration points. If a planned condition falls between two points, check how the accompanying data should be used. If you expect higher speeds or larger flow angles than those listed, make a note of those conditions and ask
WINDTUNER whether additional calibration is needed. This gives the test team time to plan for any conditions that need separate attention.
five-hole probe

Check the angle definitions before using the curves

An angle in the report only makes sense once you know which direction it refers to. When mounting the probe, compare the diagram in the report with your installation drawing. Check how the probe head faces and how the angle directions in your test log map to the ones in the report. On the curves, check the axes, units, and conditions represented. If the project includes calibration data for measurement and control software, log the data file with the probe ID from the report.

Keep the
probe calibration report with the test records. Use it before the run to check the probe, the test conditions, and the data you plan to apply. You can also refer back to it when reviewing results later. If the installation or planned conditions change, note what changed and mark the relevant pages before contacting the WINDTUNER testing center. That makes it easier to review the specific results affected.

 